Quick Search

Home / About Geneaid / Products / Q&A / Contact / ¤¤¤å / ISO 9001:2000 / Distributors / Site Map

Home Q&AWhy is the DNA recovery lower than expected?

Plasmid DNA Extraction Kit(PD004, PD100, PD300) --Q&A
¡@

Genomic DNA Extraction Kit(GT004,GT050,GT300,GB004,GB100,GB300,GP004,GP100)--Q&A
¡@

Q
Q
Why is the DNA recovery lower than expected?
A (1) Sample tissue was not lysed completely.
u Add additional Proteinase K and extend the incubation time in the Lysis step.
(2) Column was clogged at the DNA Binding Step.
u Following the Lysis Step, remove the insoluble debris by centrifugation.
u Prior to loading the column, break up the precipitate in ethanol-added lysate.
(3) Incorrect DNA Elution Step.
u Ensure that Elution Buffer was added and absorbed to the center of the GD Column matrix.
(4) Incomplete DNA Elution.
u Elute twice to increase the DNA recovery.
Q Why doesn¡¦t the Eluted DNA perform well in downstream applications?
Q Why does the genomic DNA band appear smeared in agarose gel electrophoresis?
Q What is the key to successfully isolating genomic DNA of good yield and quality?
Q Why does the sample appear viscous after sample Lysis and not pass through the column easily, even when less than the suggested maximum sample amount is used?
Q When Buffy coat is used, the column tends to be clogged.

Gel/PCR DNA Fragments Extraction Kit (DF004, DF100, DF300)--Q&A

¡@
Home / About Geneaid / Products / Q&A / Contact / ¤¤¤å / ISO 9001:2000 / Distributors / Site Map

2008 © Geneaid Biotech Ltd. All rights reserved