Plasmid DNA Extraction Kit(PD004, PD100, PD300) --Q&A
¡@
Q
Q
Why doesn¡¦t the Eluted DNA perform well in downstream applications?
A (1) Residual ethanol contamination.
     After the wash step, dry the PD Colum¬n with additional centrifugation at top speed for 5 minutes or incubate at 60¢XC for 5 minutes.
(2) RNA contamination.
u Prior to using PD1 Buffer, ensure that RNase A was added. If RNase A added PD1 Buffer is out of date, add additional RNase A.
     Too many bacterial cells were used, reduce sample volume.
(3) Genomic DNA contamination.
u Do not use overgrown bacterial cultures.During PD2 and PD3 Buffer addition, mix gently to prevent genomic DNA shearing.
(4) RNA contamination.
u Prior to using PM1 Buffer, check that RNase A was added. If RNase A added PM1 Buffer is out of date, add additional RNase A.
u Too many bacterial cells were used, reduce sample volume.
(5) Genomic DNA contamination.
u Do not use overgrown bacterial culture samples.During the PM2 and PM3 Buffer addition steps mix solution gently to prevent genomic DNA shearing.
Q Can the plasmid DNA extraction Kit be applied on Gram (+) bacteria?
Q Why do we get eluted plasmid contaminated with RNA?
Q Why can¡¦t the eluted plasmid be cut by restriction enzymes properly?
Q Why does the Plasmid DNA Extraction Kit require two washings with two types of wash buffers (W1 and Wash Buffer) while competitors kits require one washing?
Q Eluted plasmid appeared degraded after storage for an extended period of time.
Q Why was Restriction enzyme digestion unsuccessful?

Genomic DNA Extraction Kit(GT004,GT050,GT300,GB004,GB100,GB300,GP004,GP100)--Q&A
¡@

Q
Q
Q Why doesn¡¦t the Eluted DNA perform well in downstream applications?
Q Why does the genomic DNA band appear smeared in agarose gel electrophoresis?
Q What is the key to successfully isolating genomic DNA of good yield and quality?
Q Why does the sample appear viscous after sample Lysis and not pass through the column easily, even when less than the suggested maximum sample amount is used?
Q When Buffy coat is used, the column tends to be clogged.

Gel/PCR DNA Fragments Extraction Kit (DF004, DF100, DF300)--Q&A

¡@